Barocel - Capacitance manometer gauges

Capacitance manometer gauges are the most accurate “active” gauges we sell, with accuracies of <1% and the ability output a linear 0-10V signal or communicate via RS232. They are Electro-mechanical gauges using a ceramic membrane to form an electrode of a capacitor. Available in a number of temperature controlled variants, our new Capacitance range can be integrated into many applications where accuracy is key.

0.1 to 1000 Torr variants

Fast reaction time

Corrosion resistant sensor

Zero Maintenance

Fantastic zero and full scale stability

Temperature compensate

Features and benefits

• Cable connections and gauge adjustment conveniently located, thereby minimising the space envelope required for access.

 

 

 

 

 

BAROCEL Technical Data

FS options

Accuracy

Temperature effect On Zero

Admissible Temp

Ambient

Bakeout

Supply voltage (dc)

Power Consumption during heat up

%

% FS/ºC

°C

°C

V

W

7025

1000 / 100 / 10/1 / 0.1

0.2 - 0.5

0.005 - 0.02

+5 to + 50

<110

+14 to +30

<1

7025M

1000 / 100 / 10

1

0.02

+10 to +50

<110

+13 to +30

<0.3

7045

1000 / 100 / 10 / 1 / 0.1

0.15

0.0025 - 0.005

+10 to +40

<110

+10 to +40

<12

7100

1000 / 100 / 10 / 1 / 0.1

0.2 - 0.4

0.0025 - 0.005

+10 to +50

<110

+10 to +50

<15

Documentation

Spectroscopy magazine